Messages
●param.=parameterwindow
●wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
●Art1disconnectto
Art2disconnect
●Disconnected
●al.area
●param.
Noarterialinvasivepressureis
detected.
●Checkthepatientstatus.
●Checkconnections.
●Ifpressuredropsbecauseof
zeroing,performthezeroing
process.
●Art1syshigh/Art1
syslowtoArt2sys
high/Art2syslow
●Art1meanhigh/Art
1meanlowtoArt
2meanhigh/Art2
meanlow
●Art1diahigh/Art1
dialowtoArt2dia
high/Art2dialow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Artifact
●param.
IfSmartBPisenabled,this
isnormalbehaviorwhen
zeroing,ushing,orsamplingis
performed.
IfSmartBPisnotenabled,this
messageindicatesthatthe
measurementisdisturbedby
artifact.
Systolicanddiastolicnumerics
arealsoinvalidatedifanartifact
eventisdetectedwhileSmartBP
isenabled.
●Checkthepatientstatus.
●Checkcablecontact.
●Minimizetubinglength.
●Rezerothetransducer.
●Checkdevice
●ArtXpressure
malfunction
●CPPXpressure
malfunction
●CVPXpressure
malfunction
●FemXpressure
malfunction
●FemVXpressure
malfunction
●ICPXpressure
malfunction
●LAPXpressure
malfunction
●PXpressure
malfunction
●param.
●al.area
TheCARESCAPEPressurehas
malfunctioned.
●Checkthepatientstatus.
●Checkconnections.
●Iftheproblempersists,
contactqualiedservice
personnel.
220CARESCAPEONE2094480-001