Messages
param.=parameterwindow
wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
Art1disconnectto
Art2disconnect
Disconnected
al.area
param.
Noarterialinvasivepressureis
detected.
Checkthepatientstatus.
Checkconnections.
Ifpressuredropsbecauseof
zeroing,performthezeroing
process.
Art1syshigh/Art1
syslowtoArt2sys
high/Art2syslow
Art1meanhigh/Art
1meanlowtoArt
2meanhigh/Art2
meanlow
Art1diahigh/Art1
dialowtoArt2dia
high/Art2dialow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Artifact
param.
IfSmartBPisenabled,this
isnormalbehaviorwhen
zeroing,ushing,orsamplingis
performed.
IfSmartBPisnotenabled,this
messageindicatesthatthe
measurementisdisturbedby
artifact.
Systolicanddiastolicnumerics
arealsoinvalidatedifanartifact
eventisdetectedwhileSmartBP
isenabled.
Checkthepatientstatus.
Checkcablecontact.
Minimizetubinglength.
Rezerothetransducer.
Checkdevice
ArtXpressure
malfunction
CPPXpressure
malfunction
CVPXpressure
malfunction
FemXpressure
malfunction
FemVXpressure
malfunction
ICPXpressure
malfunction
LAPXpressure
malfunction
PXpressure
malfunction
param.
al.area
TheCARESCAPEPressurehas
malfunctioned.
Checkthepatientstatus.
Checkconnections.
Iftheproblempersists,
contactqualiedservice
personnel.