Messages
Forinformationregardingalarmprioritiesandescalationtimes,seethesupplemental
informationprovided.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
Analogoutput
malfunction
●al.areaCARESCAPEONEanalogoutput
voltagefailure.
●Contactqualiedservice
personnel.
●Batteryempty!
●al.areaTheCARESCAPEONEisbeing
usedonbatterypower,and
thereislessthan5minutesof
chargeleft.
●Chargethebatteryby
connectingtheCARESCAPE
ONEtoapowered
CARESCAPEDockF0.
●Batteryfailure●al.areaThebatteryisfaulty.●Replacethebattery.
●Batterylow●al.areaTheCARESCAPEONEisbeing
usedonbatterypower,and
thereislessthan20minutesof
chargeleft.
●Chargethebatteryby
connectingtheCARESCAPE
ONEtoapowered
CARESCAPEDockF0.
●Batterytemphigh●al.areaBatterytemperatureistoohigh.●Replacethebattery.
●Iftheproblempersists,
contactqualiedservice
personnel.
●CallService:Text(s)
missing
●al.areaThesoftwaretextismissingin
thislanguage;thetextlemay
becorrupted.
●Contactqualiedservice
personnel.
●Caseended●al.areaORandPACUsoftware
packages:Thereisnoactive
patientcase.
●Noactionrequired.
●Casestarted●al.areaORandPACUsoftware
packages:Anewcasehas
justbeenstarted.
●Noactionrequired.
●Conditionbattery●al.areaThebatteryneedstobe
conditioned.
●Replacethebattery,and
conditionthebatterythat
wasremoved.
●Congurationerror(s)●al.areaOneormoreerrorshavebeen
detectedintheconguration.
●Contactqualiedservice
personnel.
●DEMOMODENotfor
clinicaluse!
●al.areaDEMOmodehasbeenenabled.
●Tostartmonitoringapatient,
restartthemonitor,orselect
MonitorSetup>Defaults
&Service>ExitDEMO>
Conrm.
●Enteringstandby●al.areaActivatestandbyhasbeen
selected.
●Continuedisconnectingthe
patienttoenterstandby
mode.
230CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017