Design,environmental,andphysicalspecications
CARESCAPESpO
2
(TruSignal)
Powerrequirement-input
voltage
5VDC±0.25VDC
Powerrequirement-input
current
75mAmaximum
1mdropDropfrom1minmultipleorientationsontoahardsurface
perthetestinIEC60601-1clause15.3.4.1.
ShockWithstandsshockinaccordancewithIEC60068-27:2008
usingconditionssetinIEC60721-4-7:2001class7M3
Broad-bandrandomvibrationWithstandsbroad-bandrandomvibrationinaccordance
withIEC60068-2-64:2008usingconditionssetinIEC
60721-4-7:2001,class7M3
CARESCAPESpO
2
–Masimo
Length1.9mor1.0m(6.2or3.3ft)
Weightlessthan0.33kg(0.82lb)
Operatingtemperaturerange
0°Cto35°C(32°Fto95°F)
Non-operatingtemperature
range
-30°Cto70°C(-22°Fto158°F)
Operatinghumidityrange5%to95%RH(non-condensing)
Non-operatinghumidityrange5%to95%RH(non-condensing)
Operatingaltituderange-500mto4000m(1075hPato616hPa)
Non-operatingaltituderange-500mto5573m(1075hPato500hPa)
Degreeofenclosureprotection
againstsolidobjectsandwater
IP47
Powerconsumption2.15Wmaximum
Powerrequirement-input
voltage
5VDC±0.25VDC
Powerrequirement-input
current
430mAmaximum
1mdropDropfrom1minmultipleorientationsontoahardsurface
perthetestinIEC60601-1clause15.3.4.1
ShockWithstandshockinaccordancewithIEC60068-2-27:2008
usingconditionssetinIEC60721-4-7:2001class7M3
Broad-bandrandomvibrationWithstandbroad-bandrandomvibrationinaccordance
withIEC60068-2-64:2008usingconditionssetinIEC
60721-4-7:2001,class7M3
CARESCAPESpO
2
–Nellcor
Length3.6mor1.2m(11.8ftor3.9ft)
WeightLessthan0.20kg(0.44lb)
Operatingtemperaturerange
0°Cto35°C(41°Fto95°F)
Non-operatingtemperature
range
-40°Cto70°C(-40°Fto158°F)
256CARESCAPEONE2094480-001