Alarmspecications
AlarmLimitrangeLimitincrementAlarmdelay
MissingBeat
<5s
MultifocalPVCs
8s
Pause
6s
RonT
5s(average)(5to6srange)
STXXXhigh
whereXXX=ECGlead
label
-12.0to+12.0
mm
0.1mm
86s(average)(82to90srange)
STXXXlow
whereXXX=ECGlead
label
-12.0to+12.0
mm
0.1mm
86s(average)(85to89srange)
SVTachy
5s
Tachy20to300bpm
TheTachylimit
matchesthe
highheartrate
limit.Ifthe
highheartrate
limitischanged,
theTachylimit
changes.
1bpm
9s
Trigeminy
10s
VBrady
11s
VFib/VTach
9s
VTach
7s
VT>210s
Alarmdelayspecicationsforimpedancerespirationalarms
Thefollowingtableliststhealarmdelaysforphysiologicalalarmsrelatedtotheimpedance
respirationmeasurement.Anyalarmdelaysthatarelessthan5saregivenas<5s.
AlarmLimitrangeLimitincrementAlarmdelay
Apnea
3to30s,congurable
CardiacArtifact30breaths.Thedelaytimeisdependenton
therespirationratevalue,forexample,with
arespirationrateof60breathsperminute,
thedelayis30s.
RR(Imped.)high4to120breaths
perminute
1breathper
minute
<5s
RR(Imped.)low4to120breaths
perminute
1breathper
minute
<5s
AlarmdelayspecicationsforSpO
2
alarms
ThefollowingtableliststhealarmdelaysforphysiologicalalarmsrelatedtotheSpO
2
measurement.Anyalarmdelaysthatarelessthan5saregivenas<5s.
2094480-001CARESCAPEONE265