Parameterspecications
Theeffectsofinterferinggasesandvaporsshallbeasfollows:
ISO806061-2-55Table201.105Testgaslevelsofinterferinggasesandvaporsinnominalatmospheric
pressure,i.e.,760mmHg:
EffectofHelium(50%vol):decreasesCO
2
readings<0.5vol%
EffectofN
2
O(30%vol):increasesCO
2
readings<0.3vol%
EffectofO
2
(40%to95%):
Ifcompensationisnotactivated:decreasesCO
2
readings<0.4
vol%
Ifcompensationisactivated:CO
2
error<0.15vol%
Effectsofinterferinggasesandvapors
EffectofN
2
O(30%to80%):
Ifcompensationisnotactivated:CO
2
error<0.8vol%
Ifcompensationisactivated:CO
2
error<0.3vol%
Effectsofnon-interferinggasesand
vapors
Agasisnon-interferingifitseffecttothemeasuredCO
2
isas
follows:
0-40mmHg:±1mmHgadditionalerror
41-70mmHg:±2.5%additionalerror
71-100mmHg:±4%additionalerror
101-150mmHg:±5%additionalerror
Thefollowinggasesarenon-interferingwhentestedperISO
80601-2-55:
ethanol,
acetone,
isopropanol,
methane,
nitrogen,
andwatervapor