Defaultsettings
SettingDescription
ED
ICUNICUORPACU
Saturation
Seconds
Selectthe
saturationseconds
measurement
option.Nellcor
technologyonly.
10
Response
Selecttheaveraging
time.Nellcor
technologyonly.
Normal
Sensitivity
Selectthesensitivity.
MasimoSET
technologyonly.
Normal
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>SpO2>Alarmstoadjust
thesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
SpO2alarm
limits
Selectthehigh/low
alarmlimits.
OFF/90
Neuro
andInfant
proles:
OFF/94
OFF/90
Trauma
prole:
100/90
Neuro
prole:
OFF/94
Adoles-
cent,Pe-
diatric,
andInfant
proles:
OFF/92
100/88
Infant
prole:
100/92
OFF/90
NeuroandInfant
proles:OFF/94
SpO2alarmsonEnablealarmwhen
limitsareviolated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
PR(SpO2)alarm
limits
Selectthehigh/low
alarmlimits.
150/50
Neuroprole:120/50;
Pediatricprole:170/70;
Infantprole:180/80
200/90
Infant
prole:
180/80
160/40
Neuro
prole:
120/50;
Adolescent
prole:
150/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80
150/50
Neuro
prole:
120/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80
PR(SpO2)alarms
on
Enablealarmwhen
limitsareviolated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled.)
2094480-001CARESCAPEONE
357