Transportuse
CurrentpatientsettingstransferredfromtheCARESCAPEONEtothehostmonitor
ThefollowingcurrentpatientsettingsaretransferredfromtheCARESCAPEONEtothehost
monitoratconnectionwhenapatientisadmittedontheCARESCAPEONE.Ifapatientis
notadmittedontheCARESCAPEONEatconnection,thecurrentpatientsettingsarenot
transferredfromtheCARESCAPEONEtothehostmonitor.Instead,thecurrentpatient
settingsfromthehostmonitoraretransferredtotheCARESCAPEONEasdescribedinthe
tableCurrentpatientsettingstransferredfromthehostmonitortotheCARESCAPEONE.
ECG
●ECG1Lead
●ECG2Lead
●ECG3Lead
●LeadAnalysis
●STAnalysis
●VaLeadPosition
●VbLeadPosition
Impedance
respiration
●RespLead
●RespMeasurementon/off
SpO
2
None
Non-invasive
bloodpressure
●NIBPAuto
Invasiveblood
pressure
●Channellabel(iftheCARESCAPEONEchanneliszeroed)
●FilterHz(iftheCARESCAPEONEchanneliszeroed)
●IABPOn(iftheCARESCAPEONEchanneliszeroed)
Temperature
●T1Measurement
●T2Measurement
CO
2
None
CurrentpatientsettingstransferredfromthehostmonitortotheCARESCAPEONE
ECG
NOTE
WhenaCARESCAPEONEisconnectedtoahost
monitor,thehostmonitordisplaystheheartrate
readingsentbytheCARESCAPEONEuntilthehost
monitor’sECGalgorithmhassuccessfullylearned
theECGrhythm.
NOTE
CARESCAPEONEdoesnothavegroupSTalarm
settingsinstandalonemode.Whenahostmonitor
isusingthegroupsettingsforSTalarmlimits,
individualSTalarmlimitsandon/offstatusesare
transferredtotheCARESCAPEONEwhenitis
disconnectedfromthehostmonitorfortransport.
Uponreconnectionaftertransport,thehost
monitorgroupsettingsareusedagain.
YoushouldalwayschecktheSTalarmsettings
ontheCARESCAPEONEafterdisconnectingfor
transporttomakesuretheyareappropriateforthe
patientbeingtransported.