Transportuse
CurrentpatientsettingstransferredfromtheCARESCAPEONEtothehostmonitor
ThefollowingcurrentpatientsettingsaretransferredfromtheCARESCAPEONEtothehost
monitoratconnectionwhenapatientisadmittedontheCARESCAPEONE.Ifapatientis
notadmittedontheCARESCAPEONEatconnection,thecurrentpatientsettingsarenot
transferredfromtheCARESCAPEONEtothehostmonitor.Instead,thecurrentpatient
settingsfromthehostmonitoraretransferredtotheCARESCAPEONEasdescribedinthe
tableCurrentpatientsettingstransferredfromthehostmonitortotheCARESCAPEONE.
ECG
ECG1Lead
ECG2Lead
ECG3Lead
LeadAnalysis
STAnalysis
VaLeadPosition
VbLeadPosition
Impedance
respiration
RespLead
RespMeasurementon/off
SpO
2
None
Non-invasive
bloodpressure
NIBPAuto
Invasiveblood
pressure
Channellabel(iftheCARESCAPEONEchanneliszeroed)
FilterHz(iftheCARESCAPEONEchanneliszeroed)
IABPOn(iftheCARESCAPEONEchanneliszeroed)
Temperature
T1Measurement
T2Measurement
CO
2
None
CurrentpatientsettingstransferredfromthehostmonitortotheCARESCAPEONE
ECG
NOTE
WhenaCARESCAPEONEisconnectedtoahost
monitor,thehostmonitordisplaystheheartrate
readingsentbytheCARESCAPEONEuntilthehost
monitor’sECGalgorithmhassuccessfullylearned
theECGrhythm.
NOTE
CARESCAPEONEdoesnothavegroupSTalarm
settingsinstandalonemode.Whenahostmonitor
isusingthegroupsettingsforSTalarmlimits,
individualSTalarmlimitsandon/offstatusesare
transferredtotheCARESCAPEONEwhenitis
disconnectedfromthehostmonitorfortransport.
Uponreconnectionaftertransport,thehost
monitorgroupsettingsareusedagain.
YoushouldalwayschecktheSTalarmsettings
ontheCARESCAPEONEafterdisconnectingfor
transporttomakesuretheyareappropriateforthe
patientbeingtransported.