P.C/F.S
The AUTO test will automatically
continue when the step is judged
PASS. The AUTO test will be
immediately stopped when step
is judged FAIL.
P.C/F.C
The AUTO test will automatically
continue when the step is judged
PASS. The AUTO test will
automatically continue although
the step is judged FAIL.
0.1 ~ 999.9 s
The step will be held for specified
seconds (0.1 ~ 999.9s) until the
next step, regardless of PASS or
FAIL judgment.
P.C/F.C
Literally it works identically with
the P.C/F.C setting above.
However, this setting works on
the basis of PREVOLTAGE
framework. Refer to the section
below on page 119 for details of
PREVOLTAGE.
P.C/F.S
Literally it works identically with
the P.C/F.S setting above.
However, this setting works on
the basis of PREVOLTAGE
framework. Refer to the section
below on page 119 for details of
PREVOLTAGE.