) within an AUTO test will be
executed in a continuous manner without voltage
zeroing. See the conditions below:
1. The PREVOLTAGE of P.C/F.C and P.C/F.S
options apply to ACW, DCW and IR test
modes only. And only the same mode in
continuous MANU steps can constitute the
PREVOLTAGE test. For example, the step 1
is IR mode and the step 2 is IR mode as well.
2. When applying INIT VOLTAGE function to
several MANU steps within an AUTO test,
only the 1st MANU step will be executed.
3. The contact status should be identical among
each MANU step. When CONTACT CHK is ON
for several MANU steps within an AUTO test,
only the 1st MANU step will be executed.
4. The set voltage of following MANU step
should be greater than or equal to the
previous MANU step. If not, the set voltage of
next MANU step will only start after zeroing.
5. The frequency setting should be identical
among each MANU step within an AUTO test.
6. When applying RAMP DOWN function to
several MANU steps within an AUTO test,
only the last MANU step will be executed.
7. When applying CON function to a MANU step
within an AUTO test, it is available to
interconnect with next AUTO test page for test
based on PREVOLTAGE. However, if the 1st
MANU step is selected CON in the next AUTO
test, the PREVOLTAGE setting will be invalid.
8. If a MANU step is set SKIP within an AUTO test,
the following MANU step will not be executed
under PREVOLTAGE setting. For example, when
MANU step 2 is set SKIP, the AUTO test will start
from MANU step 1 under PREVOLTAGE followed
by MANU step 3 without PREVOLTAGE setting.