OPERATION
167
Statistics Settings
Description
This section allows user to have a
comprehensive overview of not only total test
counts including PASS and FAIL amounts,
individually, but also the respective counts of
each test mode. More than that, user is able to
view those data from an intuitive histogram.
Steps
1. Press the SYSTEM button on the
front panel when the tester is
under READY status in either
MANU or AUTO test.
2. The SYSTEM page will be shown.
Press the UP/DOWN arrow soft-
keys to move the cursor to the
STATISTICS setting where PASS
and FAIL amounts and TOTAL
amounts to date are shown in the
green highlight below. Also, the
detailed distributions of PASS and
FAIL amounts from each test
functions are well disclosed for
viewing in the blue highlight below.
PASS, FAIL amounts & TOTAL amounts
PASS & FAIL amounts
distributions in each test function