Page #212 background imageLoading...
Page #212 background image
GPT-10000 Series User Manual
212
Status - HFAIL
Applied Function:
ACW, DCW, IR, GB, CON
Description:
Read Data > HI SET
Status – LFAIL
Applied Function:
ACW, DCW, IR, GB, CON
Description:
Read Data < LOW SET
Status - VOVER
Applied Function:
ACW, DCW, IR, GB, CON
Description:
ACW, DCW, IR: Read V > V set 110%
GB: GBV > 7.2V
CON: Read CONT V > 8V
Status - V LOW
Applied Function:
ACW, DCW, IR
Description:
Read V < V set 90%
Status - SHORT
Applied Function:
ACW, DCW, IR
Description:
Read V < 10V
Contarct Check LOW Error
Status - IOVER
Applied Function:
GB
Description:
Read I > I SET 110%

Table of Contents