REMOTE CONTROL
213
Status - I LOW
Applied Function:
GB
Description:
Read I < I SET 90%
Status - ARC
Applied Function:
ACW, DCW
Description:
Read T > ARC Set Current
Status - GFAIL
Applied Function:
ACW, DCW, IR
Description:
GFCI ERROR
Status - OPEN
Applied Function:
ACW, DCW, IR
Description:
Contact Check LOW Error
Status - HOLDP
Applied Function:
ACW, DCW, IR, GB, CON
Description:
When the PASS HOLD is set for
AUTO mode, the HOLDP status
will be returned after executing
MEAS command.
Status - HOLDF
Applied Function:
ACW, DCW, IR, GB, CON
Description:
When the FAIL HOLD is set for
AUTO mode, the HOLDP status
will be returned after executing
MEAS command.
Status - HOLDT
Applied Function:
ACW, DCW, IR, GB, CON
Description:
When HOLD TIME is kept
running on, the HOLDT status
will be returned after executing
MEAS command.