tester to match the specification. After warm-up, please proceed to
Zero Check procedure. Refer to page 71 for details.
The measured value of Continuity test does not match the
specification?
Make sure the tester is powered on for warm-up of at least 30
minutes, within +15°C~+35°C. This is necessary to stabilize the
tester to match the specification. After warm-up, please proceed to
Zero Check procedure. Refer to page 71 for details.