FAQ
265
When I press the START button the tester will not start
testing?
The tester must first be in the READY status before a test can be
started. Ensure the tester displays READY before pressing the
START button, page 92 (manual test), 120 (automatic test).
If Double Action is enabled, the START button must be pressed
within 0.5 seconds after the STOP button is pressed, otherwise the
tester will not start testing.
If Interlock is enabled, the interlock key must be inserted into the
signal I/O port on the rear before a test starts. See page 186.
Lastly, if Start Click For 1 Second is enabled, it is required to press
and hold the START button for 1 second above to initiate a test.
Pressing the START button below 1 second will not start a test.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com / marketing@goodwill.com.tw.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-10000 Series and is the answer not in the manual?