Operation - Calibration
Fusion4 MSC-L Part No.: 4418309_Rev09
5 - 276 Installation & Operation Manual
Honeywell
9. On the Calibration . Stream n . Wizard 5/6 screen, select <OK> on
the IR controller or LAD, to accept the new meter factor and view the
calibration details.
The Calibration . Stream n . Wizard 6/6 screen appears, which
displays the calibration details.
A new calibration record is created and stored in the system. Each
calibration is saved in a non-volatile memory, with date, time, old and
new values.
NOTE: Select <ESC> to restore the old meter factor.