Operation - W&M Compliance
Fusion4 MSC-L Part No.: 4418309_Rev09
5 - 324 Installation & Operation Manual
Honeywell
5.26.3 Device Health
For the MSC-L to be W&M compliant the device health should be good.
The device health is good in the following scenarios:
All the hardware is available and functional.
All memory and interface checks are passed.
All boards are in valid temperature range.
All board voltages are in valid range.
If any of the board has bad or uncertain health the device health is bad.
NOTE: CAN-IN-OUT-MSC board is not taken into account
because of non relevance for W&M.
5.26.4 MSC-L W&M compliance
The MSC-L is compliant for W&M custody transfer if all the following
criteria are met.
The W&M Intended is set to <True>.
The MSC-L is W&M sealed - either the jumper in ON or the switch is
activated.
The device health is good.