GASLAB Q2 DISPLAYS (OPERATE THE DEVICE) 
Information for general use 
Rev. M / 73023639 
Action: Manual operational calibration 
Click on “Gas components” in the 
“Operational calibration” line. Check 
or correct the calibration gas mixture 
using the manufacturer’s analysis 
certificate. 
Entries and amendments ( see 
7.2.3 Entries and changes using the 
operation panel). 
 
Figure 7.52: Operational calibration 
gas composition 
Click on “Start” in the “Operational calibration” line. 
Note: If an error occurs during the following steps, for example the pressure is inadequate, the 
display will jump straight to point 8. 
The message “Calibration step failed” will be displayed. Carry out step 8.  
Rectify the error and repeat the process from step 4. 
If you are sure that the operational 
calibration gas (gas 1) is properly 
connected, click on “Confirm”.  
 
 
 
Figure 7.53: 
Confirm operational calibration gas 1 
The process will now run 
automatically, the warning 
“Operational calibration” will be 
active during this time. 
First of all, the system is flushed with 
the calibration gas mixture. This will 
be displayed throughout the duration 
of the process. 
 
Figure 7.54: 
Calibration progress for flushing with 
calibration gas  
The operational calibration gas will 
then be measured. The duration of 
this process will also be displayed. 
 
Figure 7.55: Q2 calibration progress 
for gas 1