CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
Information for general use
Rev. M / 73023639
219
GasLab Q2
8.6.6. Use “Live data tab” as Device monitor
The information about
Operating hours
Battery state of charge
CPU load
Ram load
can be obtained if the system export values are used in the tab “Live data” as shown below.
Figure 8.53: System Device monitor values
The battery is mainly used when the device is off, it must be replaced when the charge drops to
less than 20%.
A discharged battery may, in certain circumstances, result in the loss of calculated and saved
measurement data. The device will no longer start correctly.
CAUTION DANGER OF EXPLOSION IF BATTERY IS INCORRECTLY REPLACED.
The replacement of battery must be carried out by Honeywell service personnel additional a
permit from the plant operator is needed.
ATTENTION - RISQUE D'EXPLOSION SI LA BATTERIE EST MAL REMPLACÉE.
Le remplacement de la batterie doit être effectué par le personnel de service de
Honeywell et doit être approuvé par l'opérateur du système.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell GasLab Q2 and is the answer not in the manual?

Honeywell GasLab Q2 Specifications

General IconGeneral
BrandHoneywell
ModelGasLab Q2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals