6.1.1.  Set password for Administrator (admin1) 
Assign  password at  first  login to  the  enCore device!  In  the delivery state, user passwords  for 
standard users are empty for all enCore devices, including admin1. Even if you create a new user 
in enSuite, enSuite creates these users without a password. 
We strongly recommend that each user changes his or her password  
at the first log into the device. 
 
Further details in sections➪ 
2.8 Data security information, 7.3.11System display Users (Login/Logout / 
Password) and 8.3 User management and login 
 
 
6.1.2.  Adapting the factory parameterization 
In  the  factory  parameterization  (default  parameter  set),  usually  only  calibration  gas  properties, 
automatic calibration intervals and the purging / flushing times for the gases must be adjusted. For 
extreme ambient conditions, the setting for the auxiliary housing heater or the fan is still required.  
The default parameter set automatically switches on the additional heating system at 0°C ≙ 32°F 
≙ 273K with a switching hysteresis of 2°C ≙ 3,6°F ≙ 2K. This supplementary heater function is only 
available up to a maximum housing temperature of + 13°C ≙ 55°F ≙ 286K. 
Other parameters are normally left at their factory-set values and, if necessary, supplemented by 
customer-specific requirements.  
To understand these parameters and to change their values if necessary, please refer to Chapter 
 8 Configuration and analysis software enSuite and the enSuite online help.  
 
 
6.1.3.  Opening and closing the hardware parameter guard (SSW) 
You can protect settings which affect the measuring properties. This is done with the Security Switch 
(short: SSW) also called calibration switch. On delivery the connection is open. To open or close the 
SSW, the device must be opened. 
 
Never open installed devices without the permission and consent of the system operator. 
Observe the explosion protection and safety regulations! 
N'ouvrez  jamais  les  appareils  installés  sans  l'autorisation  et  le  consentement  de 
l'opérateur du système. Respectez les règles de protection contre les explosions et de 
sécurité ! 
 
Additionally, the housing hood may only be removed in controlled conditions  
as described in ISO/IEC 61010-1.