MAINTENANCE, CLEANING AND REPAIR 
Information for general use 
Rev. M / 73023639 
 
10.1.5. Checking the breather element/venting system 
Action: Monitoring the breather element 
Check whether the openings are free of dirt and blockages. 
If openings are dirty or clogged, remove the dirt.  
Do not pour or spray water or cleaning products into the breather element. 
Notify Honeywell if the breather element cannot be cleaned. 
Disconnect the gas and voltage supply. 
Do not take any further action or complete any work  
as long as the breather element is blocked. 
Note the result/condition in the maintenance report. 
 
10.1.6. Visual inspection of the connections 
Action: Visual connection inspection 
Check the connections (fluids/electrical) to determine whether they are undamaged, 
complete and free of corrosion. 
If you find any damage or corrosion, replace the damaged and/or corroded 
components and seal the unused openings. Obey the explosion protection regulations. 
If step 2 is not possible or spare parts are required, contact Honeywell to arrange a 
service visit. Disconnect the gas and voltage supply. 
Do not take any further action or complete any work until the functions have been 
restored or the defective parts have been replaced. 
Note the result/condition in the maintenance report. 
 
10.1.7. Checking the interior 
Action: Checking the interior of the device 
Disconnect the gas and voltage supply. Unscrew the connection box cover and hood 
on the instrumentation section (only in non-hazardous areas after permission of the 
plant operator 10.4 Repair work and Honeywell service work.) 
The interior is undamaged, there are no loose wires and no parts are corroded. All 
unused openings are sealed.   
If step 2 is not satisfied, contact Honeywell to arrange a service visit.  
Only reconnect the device to the power supply if step 2 is satisfied and all the covers 
have been screwed back into place. Obey the explosion protection regulations! 
Note the result/condition in the maintenance report.