COMMISSIONING AND DECOMMISSIONING 
Information for general use 
Rev. M / 73023639 
 
6.3.  Decommissioning / Dismantling / Disposal 
Please proceed in the following order if  you take the measuring instrument out of operation and may 
want to reuse it later at another location or dispose of it. 
 
Switch off the voltages (supply and communication) and secure them against being 
switched on again. Any further disassembly is prohibited without this prerequisite! 
Shut down all gas streams and depressurize all gas inlets. Prevent the unintentional 
opening of the shut-off valves by suitable measures so that no danger can arise from 
further disassembly. 
The battery buffers data for approx. one year and may need to be replaced afterwards. 
Data for example, archive data and time settings, will be lost if the battery is empty. The 
device will no longer start correctly. You will then require a Honeywell service visit. 
La batterie tamponne les données pendant environ un an et peut avoir besoin d'être remplacée par 
la suite. Les données, par exemple les données d'archives et les réglages de l'heure, seront perdues 
si la batterie est vide. L'appareil ne démarrera plus correctement. Vous devrez alors faire appel à un 
service d'entretien Honeywell. 
 
To complete dismantling or disassembly, continue the above steps with the following, note before: 
 
If the device contains hazardous media e, g. toxic gases, they must be removed before its 
dismantling. Prevent the release of potentially pollutant or dangerous substances.  
Si l'appareil contient des fluides dangereux, par exemple des gaz toxiques, ceux-ci doivent être retirés 
avant  son  démantèlement.  Empêcher  le  rejet  de  substances  potentiellement  polluantes  ou 
dangereuses. 
Action: Continue decommissioning with dismantling 
Allow the device to cool down to ambient temperature and disconnect all the gas 
tubing at the couplings. Immediately seal all outlets and inlets. 
Open the connection box in a non-potentially explosive atmosphere and remove all the 
cables by releasing them in the cable gland or cutting them if they are sealed. Dismantle 
the device mechanically. 
If you want to use the device later in another location, please bring it into the condition 
in which it was transported. Observe the instructions in section 5.1 Storage.  
Observe the following information on disposal if you no longer wish to use the device.