GASLAB Q2 AFTER DELIVERY AND AT THE PLACE OF USE 
Information for general use 
Rev. M / 73023639 
Store the GasLab Q2 in a clean, dry place. Prevent dirt getting inside the device. Prevent 
material damage caused by the formation of condensation. Gardez le GasLab Q2 dans un 
endroit propre et sec. Empêchez la saleté de pénétrer à l’intérieur de l’appareil. Évitez des 
dommages matériels dûs à la formation de condensation. 
 
The following applies to storage: 
•  The ambient conditions must be maintained. Do not expose the Q2 
temperatures below −25°C ≙ −13°F ≙ 248 K or above +55°C ≙ +131°F ≙ 328 K. 
11.4 Notes on device characteristics and operating conditions 
•  The packages may only be stored in enclosed rooms. 
•  Avoid mechanical impacts during storage. 
•  The storage time with a battery is up to one year,  
2 years if the battery is then replaced. 
 
If  the  device is  transported  in  cold  weather  or  extreme  temperature fluctuations  have occurred, 
GasLab Q2 must be slowly returned to room temperature (temperature at the place of use) in order 
to prevent damage caused by the formation of condensation. 
 
Do not remove transport guards and other safety devices, for example, for gas 
connections, before the device has been installed at its destination and if dirt can still 
penetrate it. Le retrait des dispositifs de protection de transport et des dispositif de 
protection, par exemple des raccords gaz, n’est pas autorisé avant l’installation finale de 
l’appareil, et cela demeure vrai tant et aussi longtemps que la saleté peut pénétrer 
l’appareil. 
 
Storage is also crucial for devices that are to be reused after dismantling. 
 6.3Decommissioning / Dismantling / Disposal. 
 
Removed devices should generally be stored in the same way as new devices.  
We recommend replacing the battery with a new one (for safety reasons done by 
Honeywell) if the device has not been supplied with power for more than one year. 
After two years without a power supply, the battery charge will be less than 20% and 
it must be replaced by Honeywell service personnel.  
After three years, the battery will be completely discharged. Please contact 
Honeywell if you wish to store the device for a longer period.