EasyManuals Logo

Honeywell GasLab Q2 User Manual

Honeywell GasLab Q2
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #237 background imageLoading...
Page #237 background image
MAINTENANCE, CLEANING AND REPAIR
Information for general use
Rev. M / 73023639
237
GasLab Q2
10.1.8. Visual inspection of the cylinder pressure/inlet pressures
Step
Action: Checking the calibration gas cylinder pressure
1.
Read the calibration gas cylinder pressure on the pressure gauge
2.
Note: you can empty the cylinder up to a minimum pressure of 0.5MPa 72.52psi
5bar, but the cylinder must then be replaced, regardless of the maintenance interval.
Replace the calibration gas cylinder if necessary. sections 5.3.4, 7.4.2 and 10.3.
3.
Notify Honeywell if you require a new calibration gas mixture. You can use the device
without calibration gas, but the results may not be as accurate.
4.
Note the result/condition in the maintenance report.
Step
Action: Testing the inlet pressures
1.
Read the process gas inlet pressure and the calibration gas inlet pressure on the
pressure gauges.
2.
If the process gas inlet pressure (PG) is not between 0.15 MPag 1.5 barg and 0.3
MPag 3 barg = max. inlet pressure, set the pressure to within this range. If the
calibration inlet pressure (CAL) is not between 0.15 MPag 1.5 barg and 0.25 MPag
2.5 barg = max. inlet pressure, set the pressure to within this range.
3.
If you cannot complete step 2, disconnect the gas and voltage supply. Do not take any
further action or complete any work until step 2 has been satisfied, contact Honeywell
to arrange a service visit.
4.
Note the result/setting in the maintenance report.
10.1.9. Flow rate check
Step
Action: Flow rate check
1.
Flow rates are valid (the float in the flow meter is in the area between the lower and
upper mark), a process gas pressure alarm is not emitted.
2.
If step 1 is not satisfied, adjust the flow rates. If this is not possible at the current inlet
pressure, check if the calibration gas pressure alarm is given during calibration. If no
calibration gas pressure alarm is issued, the inlet particle filter must be replaced. If an alarm is
issued in both cases, the error is in the gas management or in the sensor system.
3.
If a new adjustment cannot be made, disconnect the gas and voltage supply. Do not
take any further action or complete any work until step 1 is satisfied.
Contact Honeywell 10.4 Repair work and Honeywell service work.
4.
Note the result/setting in the maintenance report.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell GasLab Q2 and is the answer not in the manual?

Honeywell GasLab Q2 Specifications

General IconGeneral
BrandHoneywell
ModelGasLab Q2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals