GASLAB Q2 DISPLAYS (OPERATE THE DEVICE) 
Information for general use 
Rev. M / 73023639 
Action: Manual verification 
The measurement will now run for 
the time previously set in enSuite  
(see section 8). 
You can end the verification 
prematurely by selecting “Stop”. You 
will then jump to step 9. 
After the end of the time or if there is 
no test gas pressure, you will be 
requested to return to analysis 
mode.  
Select “Start analysis”. 
 
Figure 7.62: 
Display for the end of Q2 verification 
You will be requested to restore the 
original gas connection.  
(The message “Prepare process gas” 
will appear depending on the actual 
input used.) 
Figure 7.63: 
Q2 – Confirm (process gas) display 
Restore the original connection correctly. 
Obey all the safety information for working on gas connections. 
Conduct a tightness test. 
 10.1.10 System tightness test 
After you have connected the gas, accept the prompt from step 9  
(see figure above) using “Confirm”.  
The device will revert to analysis mode, and the verification is ended.