Establish the data connection to the appropriate device
( see section 8.2.4 Establish a connection to the enCore device).
The parameter set will be transferred.
Figure 8.29:
Transfer parameterization
There are typically a number of parameters which are protected by the security switch (SSW), e.g. in
order to meet the requirements of an approval. If the new device parameterization is likely to bring
about changes to such parameters, but the security switch is closed, then the message Device
could not be parameterized appears following the transfer.
The device does not restart however; the old parameterization is still used.
If the parameters are not protected by the SSW or if this is open, the changes will be transferred, and
the device will restart.
Figure 8.30: Parameterization transferred successfully
The restart will disconnect the device and the user will be logged out automatically.
The device will now operate using the transferred settings after you have re-enabled the required
operating mode. These new settings will remain in use even after the device is switched off and on
again.