MAINTENANCE, CLEANING AND REPAIR
Information for general use
Rev. M / 73023639
233
GasLab Q2
Main areas for the inspection
Check the gas connections using a tightness test with gas detector
( 10.1.10 System tightness test)
Check the battery capacity*
Check the housing parts
Check the external elements with tools (is everything secure)
Monitor the vent/breather element
Conduct a visual inspection of the connections (electrical/mechanical)
Inspect the interior of the device
Conduct a visual inspection of the cylinder pressure
Conduct a visual inspection of the inlet pressures
Adjust the flow volumes using the rotameter
Conduct a visual inspection of the integrated particle filter in the process gas
pipeline and replace it if necessary (only possible during a service visit).
Check the mechanical strength of the electrical grounding and test it for low
resistance
* If you find during inspections that a replacement is necessary, it must be done as part of a service visit,
see section 10.4 Repair work and Honeywell service work.
Replacing the consumables *
Particle filter for process gases, depending on impurities, every 1 to 3 years
Sensor block due to ageing of the bulb in the IR sensor, every 3 to 5 years
Lithium cell battery, depending on use, as a precaution every 10 years
* The replacement of the above consumables is only possible during a service visit.
If you operate the device official (fiscal), please also observe the applicable
regulations regarding sealing and closing off the housing before opening it.
If you have any questions, please contact Honeywell.