Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 4: Theory of operation
2651A-901-01 Rev. A / March 2011 4-19
Figure 81: Comparison of unguarded and guarded measurements
Insulator
I
D
DUT
Metal Mounting Plate
IL
RL1 RL2
Insulator
I
M = ID + IL
IM = Measured current
I
D = DUT current
IL = Leakage current
IN/OUT
HI
IN/OUT
LO
A: Unguarded
ID
DUT
Metal Mounting Plate
RL1
Insulator
I
M = ID
IN/OUT
HI
IN/OUT
LO
x1
GUARD
0V
SourceMeter
SourceMeter
B: Guarded
Connect to earth safety ground
using #18 AWG wire or larger.
Cable Shield
Safety Shield
I-meter
A
I-meter
A
V-source
+
-
V-source
+
-