Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-351
Usage
tsplink.trigger[N].assert()
N
The trigger line (1 to 3)
Details
The set pulse width determines how long the trigger is asserted.
Example
tsplink.trigger[2].assert()
Asserts trigger on trigger line 2.
Also see
tsplink.trigger[N].clear() (on page 7-351)
tsplink.trigger[N].mode
(on page 7-352)
tsplink.trigger[N].overrun
(on page 7-354)
tsplink.trigger[N].pulsewidth
(on page 7-355)
tsplink.trigger[N].release()
(on page 7-355)
tsplink.trigger[N].stimulus
(on page 7-357)
tsplink.trigger[N].wait()
(on page 7-358)
tsplink.trigger[N].clear()
This function clears the event detector for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
tsplink.trigger[N].clear()
N
The trigger line (1 to 3)
Details
The event detector for a trigger recalls if a trigger event has been detected since the last
tsplink.trigger[N].wait() call. This function clears a trigger event detector, discards the previous history
of the trigger line, and clears the tsplink.trigger[N].overrun attribute.
Example
tsplink.trigger[2].clear()
Clears trigger event on synchronization Line 2.
Also see
tsplink.trigger[N].mode (on page 7-352)
tsplink.trigger[N].overrun
(on page 7-354)
tsplink.trigger[N].release()
(on page 7-355)
tsplink.trigger[N].stimulus
(on page 7-357)
tsplink.trigger[N].wait()
(on page 7-358)