Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-355
Also see
tsplink.trigger[N].assert() (on page 7-350)
tsplink.trigger[N].clear()
(on page 7-351)
tsplink.trigger[N].mode
(on page 7-352)
tsplink.trigger[N].release()
(on page 7-355)
tsplink.trigger[N].reset() (on page 7-356)
tsplink.trigger[N].stimulus
(on page 7-357)
tsplink.trigger[N].wait()
(on page 7-358)
tsplink.trigger[N].pulsewidth
This attribute the length of time that the trigger line is asserted for output triggers.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
TSP-Link trigger N reset
Not saved 10-e6
Usage
width = tsplink.trigger[N].pulsewidth
tsplink.trigger[N].pulsewidth = width
width
The pulse width (in seconds)
N
The trigger line (1 to 3)
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to 20 μs.
Also see
tsplink.trigger[N].release() (on page 7-355)
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
tsplink.trigger[N].release()
N
The trigger line (1 to 3)