Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-334 2651A-901-01 Rev. A / March 2011
Details
This function sets the blender event detector to the undetected state and resets the event detector's overrun
indicator.
Also see
None
trigger.blender[N].EVENT_ID
This constant contains the trigger blender event number.
Type TSP-Link accessible
Affected by Where saved Default value
Constant Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
Trigger event number
N
The blender number (1 to 4)
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to trigger events
from this trigger blender.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender operates in OR mode or AND mode.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Trigger blender N reset
Not saved false (AND mode)
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
The orenable mode:
true: OR mode
false: AND mode
N
The trigger blender (1 to 4)
Details
This attribute selects whether the blender waits for any one event (the “OR” mode) or waits for all selected
events (the “AND” mode) before signaling an output event.

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals