Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-307
Also see
Questionable Status Registers (on page E-21)
status.questionable.*
(on page 7-300)
status.questionable.unstable_output.*
This attribute contains the questionable status unstable output summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute - - - - - - - -
.condition (R) Yes Not applicable Not saved Not applicable
.enable (RW) Yes Status reset Not saved 0
.event (R) Yes Status reset Not saved 0
.ntr (RW) Yes Status reset Not saved 0
.ptr (RW) Yes Status reset Not saved 2 (All bits set)
Usage
questionableRegister = status.questionable.unstable_output.condition
questionableRegister = status.questionable.unstable_output.enable
questionableRegister = status.questionable.unstable_output.event
questionableRegister = status.questionable.unstable_output.ntr
questionableRegister = status.questionable.unstable_output.ptr
status.questionable.unstable_output.enable = questionableRegister
status.questionable.unstable_output.ntr = questionableRegister
status.questionable.unstable_output.ptr = questionableRegister
operationRegister
The questionable status unstable output summary register's status. A
zero (0) indicates no bits set (also send 0 to clear all bits); the only valid value other
than zero (0) is 2