Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-170 2651A-901-01 Rev. A / March 2011
Example
setup.poweron = 0
Set the instrument to use the factory default setup
when power is turned on.
Also see
Start-up (power-on) configuration (on page 2-41)
setup.recall()
This function recalls settings from a saved setup.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
setup.recall(id)
id
An integer or string specifying the location of the setup to recall:
• Factory default setup (0)
• User-saved setup in nonvolatile memory (1 to 5)
• User-saved setup on a USB flash drive ("/path/filename")
Details
When the id parameter is an integer (n), it is interpreted as the setup number to restore from the instrument's
nonvolatile memory. When n = 0, the unit recalls the factory default setup; when n = 1 to 5, the instrument
recalls a user-saved setup from nonvolatile memory.
When the id parameter is a string, it is interpreted as the path and file name of the setup to restore from a file on
a USB flash drive. The path may be absolute or relative to the current working directory.
Before a setup is recalled an instrument reset is performed.
Example 1
setup.recall(2)
Recall the user-saved setup at location 2.
Example 2
setup.recall("/usb1/KEITHLEY_30730.set")
Recall a user-saved setup stored in a file
named KEITHLEY_30730 on a USB flash
drive.
Also see
User setup (on page 2-39)