Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-339
trigger.EVENT_ID
This constant contains the command interface trigger event number.
Type TSP-Link accessible
Affected by Where saved Default value
Constant Yes
Usage
eventID = trigger.EVENT_ID
eventID
The command interface trigger event number
Details
You can set the stimulus of any trigger event detector to the value of this constant to have it respond to
command interface trigger events.
Also see
None
trigger.timer[N].clear()
This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.timer[N].clear()
N
Trigger timer number to clear (1 to 4)
Details
This function sets the timer event detector to the undetected state and resets the overrun indicator.
Also see
trigger.timer[N].count (on page 7-339)
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer is triggered.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Trigger timer N reset
Not saved 1