Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-293
Details
These attributes are used to read or write to the operation status measuring summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to Status register sets
(on page E-
2) and Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings:
Bit Value Description
B0 Not used Not applicable.
B1
status.operation.measuring.SMUA
Bit will be set when taking an overlapped
measurement, but it will not set when taking a
normal synchronous measurement.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2-B15 Not used Not applicable.
In addition to the above constant, operationRegister can be set to the decimal equivalent of the bit to set.
Example
status.operation.measuring.enable =
status.operation.measuring.SMUA
Sets the SMUA bit of the operation
status measuring summary enable
register using a constant.
Also see
Operation Status Registers (on page E-20)
status.operation.*
(on page 7-266)
status.operation.remote.*
This attribute contains the operation status remote summary register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute - - - - - - - -
.condition (R) Yes Not applicable Not saved Not applicable
.enable (RW) Yes Status reset Not saved 0
.event (R) Yes Status reset Not saved 0
.ntr (RW) Yes Status reset Not saved 0
.ptr (RW) Yes Status reset Not saved 2050 (All bits set)