Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-340 2651A-901-01 Rev. A / March 2011
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
count
Number of times to repeat the trigger
N
A trigger timer value from 1 to 4
Details
If count is set to a number greater than 1, the timer automatically starts the next delay at expiration of the
previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
Example
print(trigger.timer[1].count)
Read retrigger count for timer number 1.
Also see
trigger.timer[N].clear() (on page 7-339)
trigger.timer[N].reset() (on page 7-343)
trigger.timer[N].delay
This attribute sets and reads the timer delay.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Trigger timer N reset
Not saved 10e-6
Usage
interval = trigger.timer[N].delay
trigger.timer[N].delay = interval
interval
Delay interval in seconds
N
Trigger timer number (1 to 4)
Details
Each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[N].delaylist = {interval}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
Example
trigger.timer[1].delay = 50e-6
Set the trigger timer 1 to delay for
50 µs.

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals