Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-194 2651A-901-01 Rev. A / March 2011
Also see
smuX.cal.lock() (on page 7-177)
smuX.cal.restore()
(on page 7-179)
smuX.cal.save()
(on page 7-180)
smuX.cal.unlock()
(on page 7-181)
smuX.source.calibrateY() (on page 7-211)
Adjustment
(on page B-16)
smuX.measure.count
This attribute sets the number of measurements performed when a measurement is requested.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Recall
Saved setups 1
Usage
count = smuX.measure.count
smuX.measure.count = count
count
Number of measurements.
X
SMU channel (for example, smua.measure.count applies to SMU Channel A).
Details
This attribute controls the number of measurements taken any time a measurement is requested. When using a
reading buffer with a measure command, this attribute also controls the number of readings to be stored.
If smuX.measure.count is set to a value greater than 1, any measurement delay set by
smuX.measure.delay will only occur before the first measurement, while the smuX.measure.interval
controls the interval between successive measurements.
Example
smua.measure.count = 10
Sets the SMU Channel A measure count to 10.
Also see
reset() (on page 7-151)
smuX.measure.delay
(on page 7-195)
smuX.measure.interval
(on page 7-199)
smuX.measure.overlappedY()
(on page 7-201)
smuX.measure.Y()
(on page 7-205)
smuX.reset()
(on page 7-209)
setup.recall()
(on page 7-170)