Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-352 2651A-901-01 Rev. A / March 2011
tsplink.trigger[N].EVENT_ID
This constant identifies the number that is used for the trigger events.
Type TSP-Link accessible
Affected by Where saved Default value
Constant Yes
Usage
eventID = tsplink.trigger[N].EVENT_ID
eventID
The trigger event number.
N
The trigger line (1 to 3)
Details
This number is used by the TSP-Link trigger line when it detects an input trigger.
Set the stimulus of any trigger event detector to the value of this constant to have it respond to trigger events
from this line.
Also see
None
tsplink.trigger[N].mode
This attribute defines the trigger operation and detection mode.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
TSP-Link trigger N reset
Not saved tsplink.TRIG_BYPASS
Usage
mode = tsplink.trigger[N].mode
tsplink.trigger[N].mode = mode
mode
The trigger mode
N
The trigger line (1 to 3)
Details
This attribute controls the mode in which the trigger event detector and the output trigger generator operate on
the given trigger line.