Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-21
bufferVar.collectsourcevalues
This attribute sets whether or not source values will be stored with the readings in the buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Not applicable Not applicable 0 (disabled)
Usage
state = bufferVar.collectsourcevalues
bufferVar.collectsourcevalues = state
state
Source value collection status; set to one of the following:
• 0: Source value collection disabled (off)
• 1: Source value collection enabled (on)
bufferVar
The reading buffer; can be a dynamically allocated buffer (user-defined), or a
dedicated reading buffer (such as smua.nvbuffer1)
Details
Assigning a value to this attribute enables or disables the storage of source values. Reading this attribute returns
the state of source value collection. This value can only be changed with an empty buffer. Empty the buffer using
the bufferVar.clear() function.
When on, source values will be stored with readings in the buffer. This requires four extra bytes of storage per
reading. Turning on additional collection items, such as source values (this attribute) and timestamps, decreases
the capacity of a dedicated reading buffer (for example, smua.nvbuffer1), but does not change the capacity
of a user-defined dynamically allocated buffer.
This value, off or on, can only be changed when the buffer is empty. Empty the buffer using the
bufferVar.clear() function.
Example
smua.nvbuffer1.collectsourcevalues = 1
Include source values with readings for
dedicated reading buffer 1.
Also see
bufferVar.clear() (on page 7-19)
smuX.measure.overlappedY()
(on page 7-201)
smuX.measure.Y()
(on page 7-205)
smuX.nvbufferY
(on page 7-208)
smuX.trigger.measure.Y()
(on page 7-238)
Reading buffers
(on page 3-6)
bufferVar.collecttimestamps
This attribute sets whether or not timestamp values will be stored with the readings in the buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Not applicable Not applicable 0 (disabled)

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals