Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-235
Also see
bufferVar.clearcache() (on page 7-20)
Configuring and running sweeps
(on page 3-29)
smuX.trigger.measure.action
(on page 7-235)
smuX.trigger.measure.Y()
(on page 7-238)
smuX.trigger.source.action (on page 7-241)
smuX.trigger.source.linearY()
(on page 7-243)
smuX.trigger.source.listY()
(on page 7-244)
smuX.trigger.source.logY()
(on page 7-245)
Triggering
(on page 3-32)
smuX.trigger.measure.action
This attribute controls measurement actions during a sweep.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes SMU reset
Instrument reset
Not saved smuX.DISABLE
Usage
action = smuX.trigger.measure.action
smuX.trigger.measure.action = action
action
The sweep measure action; set to one of the following values:
0 or smuX.DISABLE: Do not make measurements during the sweep
1 or smuX.ENABLE: Make measurements during the sweep
2 or smuX.ASYNC: Make measurements during the sweep, but asynchronously
with the source area of the trigger model
X
Source-measure unit (SMU) channel (for example,
smua.trigger.measure.action applies to SMU channel A)