Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-179
Also see
reset() (on page 7-151)
smuX.cal.lock()
(on page 7-177)
smuX.cal.unlock()
(on page 7-181)
smuX.measure.calibrateY()
(on page 7-193)
smuX.reset() (on page 7-209)
smuX.source.calibrateY()
(on page 7-211)
Adjustment
(on page B-16)
smuX.cal.restore()
This function loads a stored set of calibration constants.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
smuX.cal.restore()
smuX.cal.restore(calset)
X
Source-measure unit (SMU) channel (for example, smua.cal.restore() applies
to SMU channel A).
calset
The calibration set to be loaded. Set calset to one of the following values:
0 or smuX.CALSET_NOMINAL: A set of calibration constants that are uncalibrated,
but set to nominal values to allow rudimentary functioning of the instrument.
1 or smuX.CALSET_FACTORY: The calibration constants when the instrument left
the factory.
2 or smuX.CALSET_DEFAULT: The normal calibration set.
3 or smuX.CALSET_PREVIOUS: The calibration set that was used before the last
default set was overwritten.
Details
This function will overwrite the current set of calibration constants with constants read from nonvolatile memory.
This function will be disabled until a successful call to smuX.cal.unlock() is made.
If calset is not specified, smuX.CALSET_DEFAULT will be used.
Example
smua.cal.restore()
Restores factory calibration constants for SMU channel A.
Also see
smuX.cal.lock() (on page 7-177)
smuX.cal.unlock()
(on page 7-181)
Adjustment
(on page B-16)