Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-197
Details
This attribute sets the size of the stack used for filtered measurements.
Example
smua.measure.filter.count = 10
Sets the filter count for SMU Channel A to 10.
Also see
reset() (on page 7-151)
setup.recall() (on page 7-170)
smuX.measure.filter.enable
(on page 7-197)
smuX.measure.filter.type
(on page 7-198)
smuX.reset()
(on page 7-209)
Filters
(on page 3-3)
smuX.measure.filter.enable
This attribute enables or disables filtered measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Recall
Saved setups smuX.FILTER_OFF
Usage
filterState = smuX.measure.filter.enable
smuX.measure.filter.enable = filterState
filterState
The filter status. Set to one of the following values:
0 or smuX.FILTER_OFF: Disables the filter
1 or smuX.FILTER_ON: Enables the filter
X
SMU channel (for example, smua.measure.filter.enable applies to SMU
Channel A).
Details
This attribute enables or disables the filter.
Example
smua.measure.filter.enable = 1
Enables the filter for SMU Channel A. Alternatively, the
value 1 may be replaced with smua.FILTER_ON.
Also see
reset() (on page 7-151)
setup.recall()
(on page 7-170)
smuX.measure.filter.count
(on page 7-196)
smuX.measure.filter.type
(on page 7-198)
smuX.reset()
(on page 7-209)
Filters
(on page 3-3)