smuX.measure.delayfactor
This attribute stores a multiplier to the delays used when smuX.measure.delay is set to smuX.DELAY_AUTO.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Not saved 1
Usage
delayFactor = smuX.measure.delayfactor
smuX.measure.delayfactor = delayFactor
delayFactor
The delay factor multiplier.
X
Source-measure unit (SMU) channel (for example, smua.measure.delayfactor
applies to SMU channel A)
Details
The delay factor is only applied when smuX.measure.delay = smuX.DELAY_AUTO.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic delay.
Setting this attribute to zero disables delays when smuX.measure.delay = smuX.DELAY_AUTO.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay for SMU channel A.
Also see
Measure auto delay (on page 2-66)
reset() (on page 7-151)
smuX.measure.delay
(on page 7-195)
smuX.reset()
(on page 7-209)
smuX.measure.filter.count
This attribute sets the number of measured readings required to yield one filtered measurement.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Recall
Saved setups 1
Usage
filterCount = smuX.measure.filter.count
smuX.measure.filter.count = filterCount
filterCount
The number of readings required for each filtered measurement. Set to a value
between 1 and 100.
X
SMU channel (for example, smua.measure.filter.count applies to SMU
Channel A).