Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-199
smuX.measure.highcrangedelayfactor
This attribute contains a delay multiplier that is used during range changes when High-C mode is active.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Recall
Saved setups 10
Usage
delayFactor = smuX.measure.highcrangedelayfactor
smuX.measure.highcrangedelayfactor = delayFactor
delayFactor
The delay factor. Set to a value between 1 and 99.
X
SMU channel (for example, smua.measure.highcrangedelayfactor applies
to SMU Channel A).
Details
This delay multiplier is only active when High-C mode is active.
Example
smua.measure.highcrangedelayfactor = 5
Increases the delay for SMU channel A by a factor if 5.
Also see
reset() (on page 7-151)
setup.recall()
(on page 7-170)
smuX.reset() (on page 7-209)
smuX.source.highc
(on page 7-215)
High-capacitance mode
(on page 3-65)
smuX.measure.interval
This attribute sets the interval between multiple measurements.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Recall
Saved setups 0
Usage
interval = smuX.measure.interval
smuX.measure.interval = interval
interval
The interval value (in seconds). Set to a value between 0 and 1.
X
SMU channel (for example, smua.measure.interval applies to SMU
Channel A).