Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-348 2651A-901-01 Rev. A / March 2011
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
nodeNumber
Set node to a number (1 to 64)
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until the next time tsplink.reset() is executed on any node
in the system.
Each node connected to the TSP-Link must be assigned a different node number.
Example
tsplink.node = 2
Sets the TSP-Link node to number 2.
Also see
tsplink.reset() (on page 7-349)
tsplink.state
(on page 7-350)
tsplink.readbit()
This function reads the state of a TSP-Link synchronization line.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
data = tsplink.readbit(N)
data
A custom variable that stores the state of the synchronization line
N
The trigger line (1 to 3)
Details
Returns a value of 0 if the line is low and 1 if the line is high.
Example
data = tsplink.readbit(3)
print(data)
Assume Line 3 is set high, and it is then read.
Output
1.000000e+00
Also see
tsplink.readport() (on page 7-349)
tsplink.writebit()
(on page 7-359)

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals