Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-343
trigger.timer[N].passthrough
This attribute enables or disables the timer trigger pass-through mode.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Trigger timer N reset
Not saved false (disabled)
Usage
passthrough = trigger.timer[N].passthrough
trigger.timer[N].passthrough = passthrough
passthrough
The state of pass-through mode. Set to to one of the following values:
true: Enabled
false: Disabled
N
Trigger timer number (1 to 4)
Details
When enabled, triggers are passed through immediately and initiate the delay. When disabled, a trigger only
initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
trigger.timer[N].reset() (on page 7-343)
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.timer[N].reset()
N
Trigger timer number (1 to 4)

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals