Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-336 2651A-901-01 Rev. A / March 2011
Details
The trigger.blender[N].reset() function resets the following attributes to their factory defaults:
• trigger.blender[N].orenable
• trigger.blender[N].stimulus[M]
It also clears trigger.blender[N].overrun.
Also see
trigger.blender[N].orenable (on page 7-334)
trigger.blender[N].overrun
(on page 7-335)
trigger.blender[N].stimulus[M]
(on page 7-336)
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Trigger blender N reset
Not saved 0
Usage
eventID = trigger.blender[N].stimulus[M]
trigger.blender[N].stimulus[M] = eventID
eventID
The event that triggers the blender action; see Details
N
An integer representing the trigger event blender (1 to 4)
M
An integer representing the stimulus index (1 to 4)
Details
There are four acceptors that can each select a different event. The eventID parameter can be the event ID of
any trigger event.
Use zero to disable the blender input.
The eventID parameter may be one of the existing trigger event IDs shown in the following table.
Trigger event IDs*
Event ID Event description
smuX.trigger.SWEEPING_EVENT_ID
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
smuX.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measure action