Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-49
Also see
digio.trigger[N].pulsewidth (on page 7-48)
digio.trigger[N].reset()
This function resets trigger values to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
digio.trigger[N].reset()
N
Digital I/O trigger line (1 to 14)
Details
This function resets the following attributes to factory default settings:
• digio.trigger[N].mode
• digio.trigger[N].pulsewidth
• digio.trigger[N].stimulus
It also clears digio.trigger[N].overrun.
Also see
digio.trigger[N].mode (on page 7-46)
digio.trigger[N].overrun
(on page 7-47)
digio.trigger[N].pulsewidth
(on page 7-48)
digio.trigger[N].stimulus
(on page 7-49)
digio.trigger[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Digital I/O trigger N reset
Not saved 0
Usage
triggerStimulus = digio.trigger[N].stimulus
digio.trigger[N].stimulus = triggerStimulus
triggerStimulus
The event identifier for the triggering event
N
Digital I/O trigger line (1 to 14)