Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-138 2651A-901-01 Rev. A / March 2011
Details
If you want to reset a specific instrument or a subordinate node, use the node[X].reset() command.
A local node reset includes:
• Source-measure unit (SMU) attributes affected by a SMU reset are reset
• Other settings are restored back to factory default settings
A localnode.reset() is different than a reset() because reset() resets the entire system.
Example
localnode.reset()
Resets the local node.
Also see
reset() (on page 7-151)
smuX.reset()
(on page 7-209)
localnode.revision
This attribute stores the firmware revision level.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Not applicable Not applicable
Usage
revision = localnode.revision
revision
Firmware revision level
Details
This attribute indicates the firmware revision number currently running in the instrument.
Example
print(localnode.revision)
Outputs the present revision level. Sample output:
1.0.0
Also see
localnode.description (on page 7-133)
localnode.model
(on page 7-134)
localnode.serialno
(on page 7-138)
localnode.serialno
This attribute stores the instrument's serial number.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Not applicable Not applicable

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals