Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-159
Example
for name in script.user.catalog() do
print(name)
end
Retrieve the catalog listing for user scripts.
Also see
None
scriptVar.autorun
This attribute controls the autorun state of a script.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) No Not applicable See Details See Details
Usage
scriptVar.autorun = state
state = scriptVar.autorun
scriptVar
The name of the variable that references the script
state
Whether or not the script runs automatically when powered on:
"yes" (script runs automatically)
"no" (script does not run automatically)
Details
Autorun scripts run automatically when the instrument is turned on. You can set any number of scripts to autorun.
The run order for autorun scripts is arbitrary, so make sure the run order is not important.
The default value for scriptVar.autorun depends on how the script was loaded. The default is "no" if the
script was loaded with loadscript or script.new(). It is "yes" for scripts loaded with loadandrunscript
or script.newautorun().
NOTE
Make sure to save the script in nonvolatile memory after setting the autorun attribute so that the
instrument will retain the setting.
Example
test5.autorun = "yes"
test5.save()
Assume a script named "test5" is in the runtime
environment.
The next time the instrument is turned on, "test5"
script automatically loads and runs.
Also see
None