Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-71
Also see
display.clear() (on page 7-54)
display.getcursor()
(on page 7-56)
display.gettext()
(on page 7-58)
display.screen
(on page 7-67)
display.setcursor() (on page 7-69)
display.smuX.digits
This attribute sets the display resolution of the selected measurement.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Recall
Saved setups display.DIGITS_5_5
Usage
digits = display.smuX.digits
display.smuX.digits = digits
digits
Set digits to one of the following values:
• Select 4-1/2 digit resolution (4 or display.DIGITS_4_5)
• Select 5-1/2 digit resolution (5 or display.DIGITS_5_5)
• Select 6-1/2 digit resolution (6 or display.DIGITS_6_5)
X
SMU channel (smua)
Details
This attribute sets the display resolution.
Example
display.smua.digits = display.DIGITS_5_5
Select 5-1/2 digit resolution for SMU A.
Also see
Display resolution (on page 3-72)
display.smuX.measure.func
This attribute specifies the type of measurement being displayed.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
Recall
Saved setups display.MEASURE_DCVOLTS